6. Troubleshooting Guide

Problem Solving Chart

Trouble & Countermeasure

Check the following points if there should or seem to be some machine trouble. Contact Yamato’s

Technical Service Department in case trouble is not solved in spite of countermeasures below.

Trouble

Cause

Countermeasure

Digital display does not

light up on the controller.

● Power is off

● Disconnection of power cord

● Fuse blows

● Check power source

● Connect the cord of motor and

body

● Exchange of fuse(2A)

The flask will not rotate

● Switch of controller is off

● Volume knob is at the “min”

● Disconnection of motor cable

● Incomplete set-up or fastening

of steam duct cause racing

● Something touches the flask

● Turn on the switch

● Turn the knob up

● Insert into the socket on the

controller

● Fasten the steam duct holder

●remove something that contacts

Incomplete vacuumization

● Wear and deterioration of

vacuum seal

● Direction of vacuum seal is

wrong

● Cooling condenser nut is

incompletely fastened

● Glass apparatus break

● Incomplete connection of glass

apparatus

● Leak from hose joints

● Exchange of vacuum seal

●Re-set the vacuum seal

● Re-fasten

●Exchange

● Re-set

● Put vacuum grease on

●Check, re-fasten and put vacuum

grease on joints