Installation/Assembly
Arm Jack for RE200
15. Arm jack JK 200(sold separately and exclusively used for RE200)
●
●
●
●
To set at a certain height
1. Turn the lever counter-clockwise to loosen and move up or down to desired height.
→Picture 1,2
2. When you determine the position, turn the lever clockwise and fasten firmly.
→Picture 3
3. After you turn the lever to fix, set the fix ring in-between at a certain height.
When you fix , fasten firmly by lever and fix ring. If you do not fasten tightly,
arm jack will not be able to sustain the motor and it could fall.
Do not move it up or down with force when the glass apparatus is connected.
The glass apparatus may come off which can cause damage.
(Picture 1)
(Picture 2)
(Picture 3)
●
●
●
●To change the height
1. Turn the lever counter-clockwise to loosen.
(Support the arm jack securely at this time or the motor will immediately drop downward.)
2. Then refer to the above description “To set at a certain height”.