beautypg.com

Wiring diagram, Lower unit, Upper unit – Yamato Scientific IQ821 Low Temperature Incubators User Manual

Page 44

background image

40

Wiring Diagram

1

SSR1

P1

P3

ELB

H1

FM2

FM1

X3

X1

X2-1

2

1

2

3

4

5

8

7

6

CT1

AC100V

CONT1

1
2
3

CN5

1

2

CN4

CN2

3

2

2

1

CN1

1

4

3

6

5

H2

1

18

CN2

PIO1

1

18

TB1 2

1

CN9

TH1

1

18

CN8

CN1

1

18

3

10

9

MV2

MV1

13

12

15

14

3

1

2

3

R

S

C

OVR1

XS1

C1

C2

X4

X2-2

XS1

RF1

2

1

3
4

X1

20
21

X2

22

23

28

+

-

TB2 2

1

TH1

X3-1

X3-2

11

X5

J1

3

2

1

J2

3

2

1

X4

4

DEF

TF

1

SSR2

P2

H3

FM4

FM3

X8

X6

X7-1

2

3

4

5

8

7

6

CT2

CONT2

1
2
3

CN5

1

2

CN4

CN2

3

2

2

1

CN1

1

4

3

6

5

H4

1

18

CN2

PIO2

1

18

TB1 2

1

CN9

TH2

1

18

CN8

CN1

1

18

3

10

9

MV4

MV3

13

12

15

14

3

1

2

3

R

S

C

OVR2

XS2

C3

C4

X5

X7-2

XS2

RF2

2
1

3
4

X6

X7

+

-

TB2 2

1

TH2

X8-1

X8-2

11

35

36

38

39

37

34

33

32

31

41

40

58

44

45

43

42

46

52

51

50

49

53

48

47

24

25
26
27

S

R

Symbol

Part name

Symbol

Part name

Symbol

Part name

ELB

Earth leakage breaker

X1

Relay (inner heater)

PIO1

Display circuit board

P1

Terminal block

X2

Relay (refrigerator)

CT1

Current transformer

H1 Heater

(inner)

X3

Relay
(overheating prevention
device)

OVR1 Overload

relay

H2

Heater (door)

X4

Relay (solenoid valve)

C1

Operation condenser

FM1

Fan motor (inner)

DEF

Defrost circuit board

C2

Start condenser

FM2

Fan motor (refrigerator)

TH1

Temperature sensor
(double)

XS1 Start

relay

MV1

Solenoid valve (defrost)

SSR1

Solid state relay

RF1

Refrigerator

MV2

Solenoid valve
(return duct)

CONT1 PLANAR board

-

-

Symbol

Part name

Symbol

Part name

Symbol

Part name

P2

Terminal block

X7

Relay (refrigerator)

PIO2

Display circuit board

H3 Heater

(inner)

X8

Relay
(overheating prevention
device)

CT2 Current

transformer

H4

Heater (door)

X5

Relay (solenoid valve)

OVR2

Overload relay

FM3

Fan motor (inner)

DEF

Defrost circuit board

C3

Operation condenser

FM4

Fan motor (refrigerator)

TH2

Temperature sensor
(double)

C4 Start

condenser

MV3

Solenoid valve (defrost)

SSR2

Solid state relay

XS2

Start relay

MV4

Solenoid valve
(return duct)

CONT2 PLANAR board

RF2

Refrigerator

X6

Relay (inner heater)

-

-

-

-