In the Event of Failure…

Safety Device and Error Code


This unit has an automatic diagnosis function built in the controller and safety devices independent of the
controller. The table below shows the cause and the solution method when the safety device operates.

Error Code:

When an abnormal condition occurs, an error code appears and the alarm lamp lights in the

controller, the buzzer sounds simultaneously. Record the error code and turn off the power of

device immediately.

Safety Device

Notify

Cause/Solution

Sensor trouble detection

ALARM” lamp lights on,

Er.01” appears

• Temperature sensor is broken or

disconnected.

• Make a call for service.

SSR short-circuit detection

ALARM” lamp lights on,

Er.02” appears

• Triac is in short-circuit
• Make a call for service.

Heater disconnecting detection

ALARM” lamp lights on,

Er.03” appears

• Heater is disconnected.
• Make a call for service.

Memory error

ALARM” lamp lights on,

Er.15” appears

• Failure in internal memory.
• Make a call for service.

Internal communication error

ALARM” lamp lights on,

Er.17” appears

• Failure in internal communication or

temperature inputting circuit.

• Make a call for service.

Overheating

ALARM” lamp lights on,

Er.19” appears

• Overheating prevention device is in

operation.

• Reset the power supply, and then

adjust the setting temperature of the
overheating protection device.

• If the state does not recover, make a

call for service.

Measurement temperature
error

ALARM” lamp lights on,

----” appears

• Measurement value is out of display

range.

• Make a call for service.