Trouble shooting – Yamato Scientific IQ821 Low Temperature Incubators User Manual

Page 40

background image

36

In the Event of Failure…

Trouble Shooting


Problem Possible

Cause

Solution

Earth leakage breaker failure

Replace the part.

Power switch failure

Replace the part.

The device does not start
when turning on the power
switch.

Power source failure

Connect to the appropriate
power source

Heater disconnection

Replace the part.

SSR failure

Replace the part.

Temperature does not rise.

Temperature controller failure

Replace the part.

Temperature sensor failure

Replace the part.

Temperature controller failure

Replace the part.

Clogging of condenser with dust

Clean the fin on condenser

Much frost on evaporator

Defrost

Relay failure

Replace the part.

Power source failure

Connect to the appropriate
power source

Temperature does not fall.

Refrigerator failure

Repair or replace the part.

SSR failure

Replace the part.

Heater does not stop working
when the temperature reaches
setting value.

Temperature controller failure

Replace the part.





In the case if the error other than listed above occurred, turn off the power switch and primary
power source immediately. Contact the shop of your purchase or nearest Yamato Scientific
Service Office.