Operation Method
Program Operation
Program creation
POWER
1. Turn on the power
• Turn on the power switch of the unit.
• The display on the controller lights on.
The default value is displayed for about four seconds after turning
on the power. The screen then displays the initial setting. The
current temperature in furnace, operation mode character and
setting temperature of overheating prevention device are
displayed on respective screens.
2. Select program mode/program pattern
① Press PROGRAM key.
The previous program mode is displayed on the setting
temperature screen.
One more pressing the PROGRAM key changes the indication
status of the current selected program mode into blinking.
One more pressing the PROGRAM key changed the indication
into the next program mode with blinking.
② Select the required program mode, and press the ENTER key.
• In the case of selecting "PrG1", "End" is displayed on the
measurement temperature screen, and the number of the steps
registered into the setting temperature screen is displayed with
blinking.
• In the case of selecting "PrG2", "PAt" is displayed on the
measurement temperature screen, and the pattern number is
displayed on the setting temperature screen with blinking. Select
the pattern "1" or "2" using ▼▲ key.
Pressing ENTER key displays "End" on the measurement
temperature screen, and the number of the steps registered into
the setting temperature screen is displayed with blinking.
• In the case of selecting "PrG3", select the pattern ”1”, ”2”, or “3”
with the same operation.