beautypg.com

Operation method, Program operation – Yamato Scientific IQ821 Low Temperature Incubators User Manual

Page 31

background image

27

Operation Method

Program Operation

Use program repeat function

This section explains how to register the program repeat (repeating a program pattern) in

program operation.

This section explains the registration procedure of program using

repeat function in "3. Register program" above.
The procedure sets the step number to be repeated "PS-n" and
repeating times "Pc-n"(n: input step number)
① Press the SUBMENU key in stead of the ENTER key after setting

the time in the step where the repeat operation is to be used.
This enters to the repeat function setting mode.

② The measurement temperature screen displays the character

"PS-n", which indicates the step to be repeated in the program
pattern. The measurement temperature screen indicates "PS-7"
in the example because repeat function is used at the seventh
step. The step number 1 to 7 can be input in the setting
temperature display screen. Enter the number (1 in the
example) using the "▲▼".

③ Press the SUBMENU key.

The measurement temperature screen displays the character
"Pc-n", which indicates the repeating times. Enter the value of
repeating times with the "▲▼".

④ The screen goes to that for the next step when the SUBMENU

key is pressed again.

To correct or check setting…

Correction of setting during the repeat setting mode is impossible.
To correct or check the setting, end the setting of step currently input. Press the FIXED TEMP key after
the temperature setting screen for the next step appears. The screen returns to the former one and
re-setting is possible.

Note: Correction or check should be made on the program setting screen.