Before using this unit, Caution, Requirements for installation – Yamato Scientific IQ821 Low Temperature Incubators User Manual
Page 11
7
Before Using this unit
Requirements for Installation
CAUTION!
10. Choose a correct power distribution board or receptacle
• Choose a correct power distribution board or receptacle that meets the unit’s rated electric
capacity.
Electric capacity:
IQ821: 100V AC, 20A
NOTE)
There could be the case that the unit does not run even after turning ON the power. Inspect
whether the voltage of the main power is lowered than the specified value, or whether other
device(s) uses the same power line of this unit. If the phenomena might be found, change the
power line of this unit to the other power line.
11. Before/after installing
• It may cause injure to a person if this unit falls down or moves by the earthquake and the
impact. etc..To prevent, take measures that the unit cannot fall down, and not install to busy
place.
12. Handling of power code
• Do not entangle the power cord. This will cause overheating and possibly a fire.
• Do not bend or twist the power cord, or apply excessive tension to it. This may cause a fire
and electrical shock.
• Do not lay the power cord under a desk or chair, and do not allow it to be pinched in order to
prevent it from being damaged and to avoid a fire or electrical shock.
• Keep the power cord away from any heating equipment such as a room heater. The cord's
insulation may melt and cause a fire or electrical shock.
• If the power cord becomes damaged (wiring exposed, breakage, etc.), immediately turn off the
power at the rear of this unit and shut off the main supply power. Then contact your nearest
dealer for replacement of the power cord. Leaving it may cause a fire or electrical shock.
• Connect the power plug to the receptacle which is supplied appropriate power and voltage.
13. Notes for some kind of sample
• Stainless steel is used for interior; however, it may be corroded by strong acid etc. And the
door packing made of rubber may be corroded by some kind of solvent, e.g. alkaline, oil,
halogen etc.
• Much frost on the evaporator degrades the refrigeration capability, which may cause
uncontrollability of setting temperature. Be careful, especially, to treat samples with large
water content that generate much frost. Perform the defrost operation if frost is observed
through the frost inspection window.
• The equipment with large heat load cannot be used because the temperature in furnace
increases.