38

Specification

Model IQ821

Method Forced

circulation

Temperature control
range

-10 to 50℃ (Common to upper/lower unit)

Temperature
adjustment accuracy

±0.3℃ (Refrigerator is in continuous operation, Common to upper/lower unit)

Temperature
distribution accuracy

±1.0℃ (Refrigerator is in continuous operation, Common to upper/lower unit)

Time required to reach
highest temperature

20℃ to 50℃: Approx. 25min.

Time required to reach
lowest temperature

20℃ to -10℃: Approx. 45min.

Inner furnace

Stainless steel

Outer covering

Electrical zinc plated steel plate, Epoxy and melamine resin coating

Door

Electrical zinc plated steel plate, Epoxy and melamine resin coating, With lock unit

Inner door

Tempered glass

Heat insulation material

Expanded polystyrene (CFC-free)

Refrigerator

Air-cooled and full-closed type compressor, 250W×2

Cooling medium

HFC-R404A

Defrost function

Manual defrost (manually ON/automatic OFF), Cycle defrost operation

Fan of blower

Axial fan×2

Heater

Iron-chrome wire heater 550W×2

Cable port

Inner diameter: 50mm (right surface of the upper/lower unit)

Temperature control
system

VS4 type 30 steps and 3 patterns programmable controller,

PID control by microcomputer

Setting method

Digital setting by menu keys and up/down keys

Display method

Setting temperature: Digital display by four-digit green LED,

Temperature: Digital display by four-digit red LED

Timer display

1min to 99h59min, 100 to 999.5h

Operation mode

Fixed temperature, Quick auto stop, Auto stop, Auto start, Program operation

Additional functions

Timer function, Calibration offset, 3 stages auto tuning function

Sensor

For temperature control: Platinum resistance bulb,

For overheating prevention: K-thermocouple

Safety devices

Self-diagnostic functions (Memory error, Failure of sensor, Heater disconnection,
Failure of overheating prevention device, SSR short-circuit, At error, Automatic
overheating prevention device), Key lock Function, Earth leakage breaker,
Overheating prevention device