Specification – Yamato Scientific IQ821 Low Temperature Incubators User Manual
Page 42
38
Specification
Model IQ821
Method Forced
circulation
Temperature control
range
-10 to 50℃ (Common to upper/lower unit)
Temperature
adjustment accuracy
±0.3℃ (Refrigerator is in continuous operation, Common to upper/lower unit)
Temperature
distribution accuracy
±1.0℃ (Refrigerator is in continuous operation, Common to upper/lower unit)
Time required to reach
highest temperature
20℃ to 50℃: Approx. 25min.
Time required to reach
lowest temperature
20℃ to -10℃: Approx. 45min.
Inner furnace
Stainless steel
Outer covering
Electrical zinc plated steel plate, Epoxy and melamine resin coating
Door
Electrical zinc plated steel plate, Epoxy and melamine resin coating, With lock unit
Inner door
Tempered glass
Heat insulation material
Expanded polystyrene (CFC-free)
Refrigerator
Air-cooled and full-closed type compressor, 250W×2
Cooling medium
HFC-R404A
Defrost function
Manual defrost (manually ON/automatic OFF), Cycle defrost operation
Fan of blower
Axial fan×2
Heater
Iron-chrome wire heater 550W×2
Cable port
Inner diameter: 50mm (right surface of the upper/lower unit)
Temperature control
system
VS4 type 30 steps and 3 patterns programmable controller,
PID control by microcomputer
Setting method
Digital setting by menu keys and up/down keys
Display method
Setting temperature: Digital display by four-digit green LED,
Temperature: Digital display by four-digit red LED
Timer display
1min to 99h59min, 100 to 999.5h
Operation mode
Fixed temperature, Quick auto stop, Auto stop, Auto start, Program operation
Additional functions
Timer function, Calibration offset, 3 stages auto tuning function
Sensor
For temperature control: Platinum resistance bulb,
For overheating prevention: K-thermocouple
Safety devices
Self-diagnostic functions (Memory error, Failure of sensor, Heater disconnection,
Failure of overheating prevention device, SSR short-circuit, At error, Automatic
overheating prevention device), Key lock Function, Earth leakage breaker,
Overheating prevention device