4. Operating procedures
Program repeat operation
Using the program
repeat function
This section explains how to register a program pattern to be repeated
(program repeat) during the program operation.
MEASURED TEMP.
℃
HEATER
ALARM
AUTO STOP
AUTO START
ENTER
SET
TEMP.
OVER TEMP.
PROTECTOR
FIXED TEMP.
RUN
STOP
SUB
MENU
TIMER
FIXED
TEMP.
PROGRAM
PROGRAM
Using the program repeat function
This section explains how to register a program when you want to
use the program repeat function in the middle of registering a
program described in the previous section 4.
This procedure sets the number of step to return to “PS-n” and the
number of repetition “Pc-n.” (n: step number when inputting the
repeat operation)
① Set time (t-7 in the example) for the step to be repeated (step 7
in the example) and press the SUB MENU key without pressing
the ENTER key. Now you can enter the repeat function set
mode.
② Character “PS-n” that indicates “return destination select” of a
program pattern is displayed on the MEASURED TEMP. display.
In the example, the repeat function is input in the seventh step
and PS-7
is displayed on the MEASURED TEMP.
display.
You can input a number 1 to 7 of a step to return to on the SET
TEMP. display and input the number of step (1 in the example)
using the ▼▲ keys.
③ Then press the SUB MENU key.
④ Character “Pc-n” that indicates “number of repetitions” is
displayed on the MEASURED TEMP. display. Input the
number (2 in the example) using the ▼▲ keys.
⑤ Pressing the SUB MENU key again to move the screen to the
next step. The screen moves to the Sv-8 registration screen in
the example.
When you want to
correct setting
errors or confirm
settings
No alteration is allowed in the middle of the repeat set mode.
If you want to go back to the previous step for example, to correct
setting errors or to reconfirm settings you have made, finish repeat
setting once, press the FIXED TEMP. key when the screen has
changed to the temperature set screen for the next step to return the
set screen to the previous screen and then redo repeat setting
operations.
Note: Be sure to perform these operations on the program setting
screen.
If you have any questions, ask the nearest sales office or the
customer support center.