Operating procedures, Operating procedures (auto stop operation), Procedures for auto stop operation – Yamato Scientific DVS602 Constant Temperature Drying Oven User Manual

Page 21

background image

19

4. Operating procedures

Operating procedures (auto stop operation)

Procedures for auto
stop operation

Used when you want to “set automatic stop after set time has elapsed
from the start of the fixed temperature operation.”

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM


MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

1. Setting a stop time

① Press the TIMER key in the initial screen.
② The timer mode you used in the previous session is displayed on

the SET TEMP. display.
Pressing the TIMER key again to flash the timer mode.
Pressing the TIMER key again to flash the next timer mode.

Select the character AStP

that indicates the auto stop

operation and press the ENTER key.

The character Sv

that indicate a set temperature is

displayed on the MEASURED TEMP. display on which the
current set temperature flashes and the AUTO STOP lamp
flashes.

③ Set a temperature you want using the ▼▲ keys.

④ Press the ENTER key. The character tim

that indicates

the timer is displayed on the MEASURED TEMP. display and the
current set time flashes on the SET TEMP. display.

⑤ Set a time you want using the ▼▲ keys.

About the timer

function

The maximum time that can be set for the timer is 999 hours 50
minutes.
Up to 99 hours 59 minutes, time can be set in minutes.
One hundred hours and over are set only in 10 minutes.
Keep the ▼▲ keys pressed to continuously change set time and you
can quickly reach the time you want. Press the ▼▲ keys once at a
time for fine adjustment.

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

2. Starting timer operation

When you have set a time you want, start the RUN/STOP key for
about one second.
Timer operation starts with the AUTO STOP lamp is illuminated.
Timer starts counting when the temperature in the bath (measured
temperature) reaches the set temperature.