beautypg.com

Operating procedures, Preparing a program, Request for operation check – Yamato Scientific DVS602 Constant Temperature Drying Oven User Manual

Page 28

background image

26

4. Operating procedures

Preparing a program

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

⑨ Press the ENTER key. The character t-1

that

indicates the set time of the first step is displayed and the
current set time flashes.

* To set a time you need to know the heating capacity (or

cooling capacity) of the unit beforehand.

* As an example, DVS 402 model requires about 75 minutes

to increase room temperature to 250 ℃ . Therefore,
assuming the current temperature to be 25 ℃ , heating
capacity can be estimated roughly as three minutes to raise
one ℃ meaning 75 minutes is necessary to increase to
250℃. In the example, the time has been estimated as 80
minutes with some allowances added.

* The maximum time that can be set for the timer of each step

is 999 hours 50 minutes.

⑩ When time has been set, press the ENTER key.

The temperature set character Sv-2 for the second step is

displayed.

Input temperature and time of each step following the

programming sheet.

⑪ When you want to repeat a certain program pattern (program

repeat) in the middle of a program as in the example, a special
procedure will be required. In this case, set time (t-7 in the
example) for the step to be repeated (step 7 in the example)
and press the SUB MENU key without pressing the ENTER
key. Now you can enter the repeat function set mode.

*For operations of the program repeat function and how to
register (input), follow procedures in “Program repeat
operation” on page 28.

⑫ When setting the temperature and time for the final step has

been completed, the screen returns to the initial setting screen.

Request for operation
check

Prior to start actual operation with samples installed, be sure to
perform no-load operation to check if the set temperature and the
set time is appropriate.

MEASURED TEMP.

HEATER

ALARM

AUTO STOP

AUTO START

ENTER

SET

TEMP.

OVER TEMP.

PROTECTOR

FIXED TEMP.

RUN

STOP

SUB

MENU

TIMER

FIXED
TEMP.

PROGRAM

PROGRAM

5. Starting the program operation

Press the RUN/STOP key for about one second. The set
program operation starts.
The PROGRAM lamp illuminates and steps of the present

operation will be displayed from the step St-1

on the SET

TEMP. display.
*Press the ▼ key during operation to check the set temperature

and the remaining time for each step being operated on the SET
TEMP. display.

This manual is related to the following products: