HEPA/UV3 PCR Workstation and Cabinet
Page 3
HEPA/UV
Three stage filter
UV Source
Pre-filter, carbon-filter and HEPA-filter
One 8-watt 254nm shortwave UV tube
Chamber Light Sources
Workstation:
UV Source - Chamber
UV Air Circulator
White Light
Cabinet:
UV Source - Chamber
UV Air Circulator
White Light
Two 25-watt 254nm shortwave UV tubes
One 8-watt 254nm shortwave UV tube
Two 8-watt white tubes
Four 8-watt 254nm shortwave UV tubes
One 8-watt 254nm shortwave UV tube
Two 8-watt white tubes
Power Outlets
Built-in outlets for operating instruments inside the chamber:
Four in the Workstation, two in the Cabinet
Shelves
Construction
Dim. (for Workstation)
Dim. (for Cabinet)
Two - Formed antimicrobial coated aluminum
13”W x 4.2”D (330 x 107mm)
8.75” x 4”D (222 x 102mm)
Construction
Interior
Antimicrobial coated stainless steel (horizontal work surface)
and aluminum
Exterior
Aluminum powder coated
Door and Side Panels
Two panel Makrolon
®
assembly
Blocks UV under 400nm
Electrical
115V - 60 Hz, 230V - 50 Hz, 100V - 50/60Hz
Operational Ratings
•
Unit intended to be used indoors
•
Altitude must not exceed 2000 m
•
Ambient temperature not to exceed 5°C to 40°C (41°F to 104°F)
•
Relative humidity not to exceed 80% for temperatures up to 31°C (88°F) decreasing linearly to 50%
relative humidity at 40°C (104°F)
•
Mains supply voltage fluctuations up to ±10% of the nominal voltage
•
Pollution rating 2 or better for laboratory areas
Part
Number
Voltage
Max
Amps
95-0434-01
95-0434-02
95-0434-04
95-0434-03
95-0434-05
95-0438-01
95-0438-02
95-0438-04
95-0438-03
95-0438-05
115V
230V
230V
100V
230V
115V
230V
230V
100V
230V
9.1
8.7
8.7
9.1
8.7
9.1
8.7
8.7
9.1
8.7