HEPA/UV3 PCR Workstation and Cabinet
Page 12
Maintenance, Replacement Parts/Accessories
and Technical Assistance
Replacement Parts and Accessories
Replacement parts are listed below. Only authorized UVP service personnel should perform repairs or
replacements other than specified in these procedures.
Replacement Parts
Part Number
Pre-filter
38-0314-01
Carbon-filter
76-0352-01
HEPA filter
17-0113-01
Shelf, small 8.8”W (for cabinets)
20-1209-01
Shelf, large 13”W (for workstations)
20-0970-02
Fuse (2 required)
56-0049-01
Makrolon
®
Panels
Side Panels (2 Required)
10-0618-01
Upper Door Panel (Workstation)
10-0619-01
Lower Door Panel (Workstation)
10-0620-01
Upper Door Panel (Cabinet)
10-0621-01
Lower Door Panel (Cabinet)
10-0622-01
Replacement Tubes
System/Part Number
Replacement Tube
Tube Description
Qty.
UV3 HEPA PCR Cabinet
34-0007-01
8 Watt 254nm UV
6
95-0434-xx
34-0056-01
8 Watt White
2
UV3 HEPA PCR Workstation
34-0007-01
8 Watt 254nm UV
2
95-0438-XX
34-0056-01
8 Watt White
2
34-0073-01
25 Watt 254nm UV
2
Accessories
Part Number
Table, stainless steel top
98-0077-01
UVX Radiometer
97-0015-02
UVX Sensor (UVX-25)
97-0016-01
J-225 Meter
97-0004-01
Spectacles, UV blocking (UVC-303)
98-0002-01
Goggles, UV blocking (UVC-503)
98-0002-02
Face Shield, UV blocking (UVC-803)
98-0002-04
Replacing the Filters
Recommended filter replacement schedule to maintain optimum filter performance:
HEPA filter:
Every 6 months or when significant drop in airflow is observed
Carbon filter:
Once a year
Pre-filter:
Once a year
PCR Table