1. To replace any of the filters or the UV tube in the
HEPA/UV system, unscrew knobs (2) located on
the ceiling of the workstation to release
compression in the HEPA/UV system.
2. Open the HEPA/UV door located on the left of the
workstation to access the filters.
3. Pre-filter. The Pre-filter (38-0314-01) is located on
the inside of the HEPA/UV door. To remove, open
the door and slide the filter off of the door.
4. Carbon Filter. Slide the filter assembly out of the chamber. The Carbon Filter (76-0352-01) then
slides to the right out of the filter assembly. If the filter does not slide out easily, loosen the thumb
screws in front of the filter.
5. HEPA Filter. To replace the HEPA filter, slide the filter assembly out of the chamber and
remove the carbon filter if not already removed. Turn the knobs counterclockwise to loosen.
Open the HEPA door.
Slide the HEPA filter out and replace filter. Close door and tighten the knobs (turn clockwise). Insert
the carbon filter. Tighten the carbon filter thumbscrews if loose. Close filter door. Retighten the
knobs (2) on the workstation ceiling.
Knobs
UV Tube
Pre-Filter
Carbon
Filter
HEPA
Filter
Thumb
screws