beautypg.com

Hepa, Kit installation manual, Startup and operation – Titus HEPAlert IOM for RadiaTec Diffusers User Manual

Page 5: Calibration, Maintenance

background image

HEPA

LERT

Kit Installation Manual

IOM-HEPA-RT-00

725343-01

09-30-04

5 of 7

Startup and Operation

When power is supplied, the green LED comes on,

indicating the system is functioning.
The red LED comes on when a pressure differential of

50% is reached, indicating that filter needs to be replaced.
Once a clean filter is inserted in the filter frame, the green

LED comes on.

Calibration

Each HEPAlert kit sensor must be set based on the

module and inlet size. Use the information in the following

steps to calibrate the setscrew.

1.

Turn the setscrew clockwise to ensure it is in the fully

closed position. See Figure 14.

2.

Based on the module and inlet size, twist the

setscrew counter-clockwise the prescribed number of

turns from the fully closed position. See Table 1.
Figure 15 shows the setscrew at the fully open

position.
Figure 16 is a chart showing the pressure from supply

side of the HEPA filter.

Table 1. RadiaTec-AL and RadiaTec-SS

Calibration Settings

Module and

Inlet Size

1- or 2-way

Pattern

CFM

Number of CCW Turns

from Fully Closed

Position

(50% Full Filter Setting)

200

2 ¾

300

2 ½

24 x 24

inches

8-inch inlet

400

2

300

2 ½

400

2 ¼

24 x 24

inches

8-inch inlet

500

2

400

2 ½

600

2 ¼

48 x 24

inches

10-inch inlet

800

2

600

2 ½

800

2 ¼

48 x 24

inches

12-inch inlet

1000

2

Note: Replace filter when filter is 50% full.

Figure 14. Setscrew at Fully Closed Position

Figure 15. Setscrew at Fully Open Position

Maintenance

Components of the HEPAlert require no regular

maintenance. However, if neither LED is on, check the

following items.

Power supplied to unit.

Wiring and tubing connections are secure.

Transformer is operational (use a voltmeter).

Contact a Titus representative if service or replacement is

required.