Figure 14: exhaust fitting connections – RKI Instruments Data Cal 2000 User Manual
Page 24

19 • Data Cal 2000 Hardware Setup
CAUTION: The maximum recommended length for the
exhaust tube is 10 feet. Do not use more than 10
feet of tubing for the exhaust tube or bump test and
calibration accuracy will be adversely affected. The
tube that is shipped with the Docking Station is 10
feet long.
GX-2009
The DM-2009 only has one exhaust fitting instead of two like the
DM-2001 and DM-2003. Each DM-2009 in a Data Cal 2000 system
must have a dedicated exhaust tube.
Install the 5/16” ID flexible tube that is included with the DM-2009
on the exhaust fitting port and route the tube to an area where the
calibration gas can be safely exhausted.
CAUTION: The maximum recommended length for the
exhaust tube is 10 feet. Do not use more than 10
feet of tubing for the exhaust tube or bump test and
calibration accuracy will be adversely affected. The
tube that is shipped with the Docking Station is 10
feet long.
MODULE
PPC (RS-232C)
Route to open area
where calibration gas
can be exhausted.
Tubing, User Supplied
GAS OUT
AIR
GAS 1
GAS 2
Figure 14: Exhaust Fitting Connections