Figure 13: exhaust fitting connections – RKI Instruments Data Cal 2000 User Manual
Page 23

Data Cal 2000 Hardware Setup • 18
3. Repeat step 2 for each successive module in the system until you
get to the last module.
4. Install the black fitting plug on the left CAL. GAS fitting of the last
module in the system.
Module Exhaust Fitting Connections
GX-2001 and/or GX-2003
1. Install a clear fitting plug, supplied with the system, on the bottom
port of the exhaust fitting of the first module in the system.
2. Install a clear tube, supplied with the system, between the top
exhaust fitting port on the first module and the bottom exhaust
fitting port on the next module.
3. Repeat step 2 for each successive module in the system until you
get to the last module.
4. On the last module in the system, install the 5/16” ID flexible tube
that is included with the Docking Station on the top exhaust fitting
port and route the tube to an area where the calibration gas can be
safely exhausted.
MODULE
PC(RS-232C)
GAS OUT
AIR
GAS1
GAS2
Install plug over fitting.
Tubing,Supplied with System
Tubing,
User Supplied
Route to open area
where calibration gas
can be exhausted.
Figure 13: Exhaust Fitting Connections