SC-01 Operator’s Manual

Maintenance • 32

Table 7: Troubleshooting the 01 Series

Symptoms

Probable Causes

Recommended Action

• The LCD is blank.

• The unit may have been

turned off.

• The alkaline batteries

may need to be
replaced.

1. To turn on the unit, press and

briefly hold the POWER
button until you hear a beep.

2. If the unit does not turn on,

replace the alkaline batteries.

3. If the difficulties continue,

contact RKI Instruments, Inc.
for further instruction.

• The LCD shows

abnormally high
readings but other
gas detection
instruments do
not.

• The unit may need to be

calibrated.

• The sensor may need

replacement.

1. Calibrate the unit.
2. Replace the sensor and

calibrate the unit.

3. If the difficulties continue,

contact RKI Instruments for
further instruction.

• “SPAN CAL FAIL”

displays during
span adjustment.

• The calibration value

may not match the
cylinder gas
concentration.

• The calibration tubing is

not connected to the
correct calibration cup
port.

• The sample gas is not

reaching the sensor
because of a bad tubing
connection.

• The calibration cylinder

may be out of gas or is
outdated.

• The sensor may need

replacement.

1. Check the calibration tubing

for leaks or for any bad
connections.

2. Confirm that the calibration

tubing is connected to the
correct calibration cup port.

3. Make sure the SC-01

calibration gas value has
been properly set up to match
the calibration cylinder
concentration.

4. Verify that the calibration

cylinder contains an
adequate supply of fresh test
sample.

5. If the fail condition continues,

replace the sensor.

6. If the difficulties continue,

contact RKI Instruments, Inc.
for further instruction.

• “CIRCUIT FAIL”

displays on the
LCD.

• A circuit failure has

occurred.

1. Turn off the unit and turn it on

again.

2. If difficulties continue, contact

RKI Instruments, Inc.