29 • Setup Mode
SC-01 Operator’s Manual
setpoint. The unit will proceed to the TWA Screen. The TWA setpoint is
flashing.
8. Use the AIR
▼
and
▲
DISP buttons to adjust the TWA setpoint to the
desired value. To turn the TWA alarm off, adjust the alarm point down to
towards zero and continue to adjust it down with the AIR
▼
button until
“OFF” is displayed.
9. Press and release the POWER button to accept the displayed TWA
setpoint. The following LCD will show the following screen before
returning to the ALARM-P screen.
The standard alarm point settings for the various models of the SC-01 are
listed below in Table 6.
Table 6: Standard Alarm Point Settings
Target Gas
Warning
Alarm
STEL
TWA
Ammonia (NH
3
)
25.0 ppm
35.0 ppm
35.0 ppm
25.0 ppm
Arsine (AsH
3
)
0.50 ppm
1.00 ppm
OFF
OFF
Chlorine (CL
2
)
0.50 ppm
1.00 ppm
1.00 ppm
0.50 ppm
Carbon Monoxide (CO)
25.0 ppm
50.0 ppm
OFF
25.0 ppm
Hydrogen Cyanide (HCN)
5.0 ppm
10.0 ppm
4.7 ppm
OFF
Hydrogen Sulfide (H
2
S)
10.0 ppm
20.0 ppm
15.0 ppm
10.0 ppm
Phosphine (PH
3
)
0.30 ppm
0.60 ppm
1.00 ppm
0.30 ppm
Sulphur Dioxide (SO
2
)
2.00 ppm
5.00 ppm
5.00 ppm
2.00 ppm
TWA
5.00
PPM
ALARM-P
End
PPM