SC-01 Operator’s Manual

Setup Mode • 28

NOTE: The sensor has the standard alarm points saved in its memory.

Changing the alarm points in this menu item does not affect the
alarm points that are saved in the sensor. If the sensor is moved to
another SC-01, the standard alarm points will be loaded.

1. When the ALARM-P Screen is displayed, press and release the

POWER button. The WARNING Screen appears displaying the
Warning setpoint. The setpoint is flashing.

2. Use the AIR

and

DISP buttons to adjust the Warning setpoint to

the desired value. To turn the Warn gas alarm off, adjust the alarm point
down to towards zero and continue to adjust it down with the AIR

button until “OFF” is displayed.

3. Press and release the POWER button to accept the displayed Warning

setpoint. The unit will proceed to the ALARM Screen. The Alarm
setpoint is flashing.

4. Use the AIR

and

DISP buttons to adjust the Alarm setpoint to the

desired value. To turn the Alarm gas alarm off, adjust the alarm point
down to towards zero and continue to adjust it down with the AIR

button until “OFF” is displayed.

5. Press and release the POWER button to accept the displayed Alarm

setpoint. The unit will proceed to the STEL Screen. The STEL setpoint
is flashing.

6. Use the AIR

and

DISP buttons to adjust the STEL setpoint to the

desired value. To turn the STEL alarm off, adjust the alarm point down
to towards zero and continue to adjust it down with the AIR

button

until “OFF” is displayed.

7. Press and release the POWER button to accept the displayed STEL

WARNING

2.00

PPM

ALARM

5.00

PPM

STEL

2.00

PPM