RaySafe i2 Installation and maintenance manual – Installation

Reserved slots menu

Figure 6. The reserved slots menu.

this is an optional setting, which you can use to reserve a number of slots in the online view for the clinical
roles doctor, nurse and technician. If you, for example, have reserved three slots for doctors but the real time
display detects only one, there will be two empty slots before the first nurse appears.

If you want to reserve slots for nurses, you also have to reserve slots for doctors. If you want to reserve slots for
technicians, you also have to reserve slots for doctors and nurses.

the default behavior “Auto” is first-come, first-served, which means that dosimeters will appear on the screen
in the order the real time display detects them, sorted after their clinical role: doctor, nurse, technician, other.
this is the recommended setting.