RaySafe i2 System User Manual – technical data
SeCURIty And pRIvACy ReqUIReMentS
AntI vIRUS
An AvS solution is not installed on this windows Ce device because of the security measures taken to reduce
the attack surface.
netwoRk poRtS
the following ports and protocols are open on the real time display for communication with the dose manager:
TCP/UdP
PORT NUmbER
PROTOCOL
AddITIONAL NOTE
tCp
8070
gSoAp httpd 2.7
Used for regular (bidirectional)
communication between real time display
and dose manager
Udp
8060
proprietary discovery
protocol
protocol (bidirectional) used for real time
display discovery
enCRyptIon
RaySafe i2 data is encrypted during transmission, at rest on the real time display and is stored in an encrypted
dose manager database.