transfer data to a CoMputer

to view the waveform in the raysafe Xi view and/or transfer measured data to a pC, connect a serial cable or
Bluetooth module (option) to the communication port on the raysafe Xi base unit.

Battery and CharGer

CAUTION!

Only authorized personnel may remove and replace the battery.

fully charge the battery with the provided battery charger before first use. Battery status in percentage is
displayed, together with setup information, when the instrument is switched on.

raysafe Xi/unfors Xi platinum plus

unfors Xi platinum

FOR RAySAFE Xi

the provided battery charger may be connected during all measurements.

FOR UNFORS Xi PLATINUM

the provided battery charger may be connected during measurement except when measuring mas, as it may
create ground currents affecting the measurement. however, the raysafe Xi battery charger may be connected
during all measurements.

Xi version

Battery type

Hours of operation

Charge time

RaySafe Xi

7.4 V Li-ion

20–40 hours

4 hours

Unfors Xi Platinum

9 V niMh

10–20 hours

15 hours

9