MeasureMent defInItIons

figure 1 illustrates the raysafe Xi parameter definitions available on a typical waveform. In most situations
the default settings are suitable for accurate measurements. In certain situations advanced users can change
the measuring definitions of the raysafe Xi by activating trig delay and/or trig level settings. Because the
measured parameters will be affected by the settings of delays and trig levels, it is recommended to note the
setup information shown on the raysafe Xi base unit display at power on.

Figure 1. Normal Measurement

When an exposure exceeds 4 seconds the measured values are continuously updated in the display. It
may take another few seconds to get stable readings (synchronized with the frame rate).