for positioning of the raysafe Xi Ct detector free in air, unfors raysafe recommends the use of the raysafe
flexi stand.

how to Measure Ct kVp

position the raysafe Xi r/f detector along the Ct beam slice. turn on the meter and select r/f high. expose
without rotation i.e. in scout/tomogram/topogram mode, using no or slowest table movement.

1. SENSOR MENU

CT Dose

Ion chamber sensor

press STEP to step between sensors and SELECT to select a sensor.

NOTE!

When turning on the RaySafe Xi base unit with the RaySafe Xi CT detector connected,
“Stabilizing” will be displayed for a few seconds while the electronics are stabilizing.