Mammography measurement, Special measurement cases – RaySafe Xi User Manual
Page 14
![background image](/manuals/571016/14/background.png)
MaMMoGraphy MeasureMent
place the detector along the X-ray tube axis (with the cable end pointing towards the X-ray machine). for kV
measurements, make sure that the center of the MaM sensor square is placed 6 cm from the examination table
front edge. dose and hVL can be measured from other positions as well.
the following concerns unfors Xi platinum, not raysafe Xi:
NOTE!
When measuring on mammography machines with carbon fibre table, we recommend not to
use the power adapter to charge Xi, or use something that creates at least a 5 mm distance
between the table and the detector, such as a plastic plate.
speCIaL MeasureMent Cases
the table below indicates when to use filter for kV measurements.
Beam quality
Filter
Comment
Mo/Mo
no filter
Mo/rh
2 mm al
w/rh
no filter
w/al 20–39 kV
no filter
w/al 40–49kV
2 mm al
use the r/f high detector
w/alscanning
2 mm al
Mo/Rh
when measuring kV with Mo/Rh, use the included 2 mm al filter. follow the instructions printed on the filter.
14
raysafe Xi user Manual – Mammography Measurement