r/f MeasureMent

for best accuracy, center the selected sensor field (r/f low or r/f high) and position the long axis of the sensor
field perpendicular to the anode-cathode axis of the tube.

1. SENSOR MENU

R/F low

sensor for conventional low dose rate measurements lower than 1 mGy/s (7 r/min),
normally after a phantom.

R/F high

sensor for conventional high dose rate measurements higher than 1 mGy/s (7 r/min),
normally before a phantom.

press STEP to step between sensors and SELECT to select a sensor.

2. MEASURE MODE

the displayed values are updated after each exposure, or continuously after 4 seconds of fluoroscopy.

kVp

kVp (for the r/f high sensor at adequate signal levels) or kV

Dose

Gy (air kerma, free in air) or r (exposure)

Dose rate

Gy or r per s, m or h

Time

ms, s

HVL

mm al

Total Filtration, TF mm al