Vu + ppm – NTi Audio Minilyzer ML1 User Manual
Page 28

28
vu + PPM
ML1 provides vu + PPM (Peak Program Meter) measurement results,
following IEC60268.
Reference
Level
Numerical
Peak-Hold
Graphical
Peak-Hold
Reference Indicator
Integration Time
Fig 3.11 vu + PPM (Type I) Panel
The PPM may be measured according to three standards:
•
Type I
•
Type IIA
•
Nordic standard
One of these standards has to be selected when entering the vu+PPM
mode through the measurement functions menu.
The following chapters describe the different entry fields and indicators
of the vu+PPM display.
a.
Reference Level & Indicator
The reference level indicates the RMS level of a 1 kHz sine wave that
defines the value of the reference indicator. The reference level may
be defined by the user in a range of -20 dBu to +14 dBu.
The default reference level depends on the selected standard
•
+4 dBu for the vu-meter
•
+6 dBu for the PPM standards Type I and Nordic
•
+8 dBu for the PPM standard Type IIA.