Integrated operations platform specifications, Integrated operations platform specifications 16 – MTS Criterion Series 60 User Manual
Page 16
![background image](/manuals/568068/16/background.png)
MTS Criterion™ Series 60
16
System Specifications
Site Preparation
Integrated Operations Platform Specifications
The following tables provide the specifications for the Integrated Operations
Platform.
General specifications
Model-specific
specifications
P
ARAMETER
S
PECIFICATION
Environmental
Operating temperature
Humidity
Noise
rating at 1 m
For use in a controlled environment
5–40°C (41–104°F)
5–85% noncondensing
70 dB (A) fully compensated
Hydraulic fluid
Reservoir capacity
Exxon Mobil DTE 25
85 L (22.5 gal) maximum
Electrical power
Line voltage
Control voltage
3~3W+PE, 400 V 50 Hz or 480 V 60 Hz
24 V DC
Dimensions
Height
Length
Width
1900 mm (75 in)
1040 mm (41 in)
720 mm (28 in)
P
ARAMETER
64.305 /64.305EL S
PECIFICATION
64.605/64.106/64.605EL/64.106EL
S
PECIFICATION
Pump/motor
Line voltage starter configuration
Line voltage starter configuration
Maximum continuous
pressure
23 MPa (3300 psi)
23 MPa (3300 psi)
Maximum flow capacity
3.3 L/m (0.9 gpm) at 50 Hz
4.0 L/m (1.0 gpm) at 60 Hz
5.1 L/m (1.3 gpm) at 50 Hz
6.2 L/m (1.6 gpm) at 60 Hz
Motor rating
1.5 kW (1.6hp) at 50/60 Hz
2.2 kW (2.9 hp) at 50/60 Hz
Heat load (maximum)
12.3 kW (42,000 Btu/hr)
20.5 kW (70,000 Btu/hr)
Weight with maximum oil
85 L (22.1 gal)
480 kg (1058 lb)
480 kg (1058 lb)