MTS SWIFT 50 GLP Sensor Heavy Trucks User Manual
Page 79

SWIFT 50 GLP Sensors
Troubleshooting
79
Errors reported when 
TI2STATUS is run.
Shunt errors that 
TI2STATUS could report.
•
Reference Bad – a downloaded shunt 
reference is zero. This is a calibration 
settings problem.
•
Shunted Bad – A bridge being shunted 
deviated from the shunt reference by more 
than the limit. This could be caused by a 
sensor failure, transducer cable failure, TI 
failure, bad shunt reference setting or bad 
shunt tolerance setting.
•
Unshunted Bad – A bridge not being 
shunted deviated by more than the limit. 
This is likely caused by a short in the 
transducer or transducer cable.
Zero errors that 
TI2STATUS could report.
•
Bad Angle Offset – the transducer should 
have been rotated +/-90 degrees but the 
measured angle was different by more than 
the limit (currently 2 degrees). This could 
be caused by not passing the index mark 
before starting the zero, not rotating by the 
correct angle, a bad angle sensor or a bad 
transducer cable.
•
Direction Change – the direction the 
transducer was rotated changed during the 
zero process. The transducer needs to be 
rotated in 90 degree increments the same 
direction in each step of the zero process so 
data can be collected for each one of the 
sensor beams (0, 90, 180 and 270 degrees).
•
Missing Data Points – a zero method that 
collects data while the transducer is rotated 
found too many consecutive missing data 
points. This is likely caused by rotating the 
transducer too fast.
•
Invalid Algorithm – the zero algorithm 
specified is not recognized. This is likely 
the calibration setting problem that could 
be affected by software and firmware 
version.
Troubleshooting Guide (part 8 of 11)
Symptom
Possible Causes
Solution
