SWIFT 50 GLP Sensors
Troubleshooting
75
The AngleOffset value is 
incorrect in the calibration 
file.
If an incorrect AngleOffset value is set in the TI 
calibration file, an axis which should have no 
load may show some load from another vehicle 
coordinate axis based on the incorrect reference 
orientation. Verify that the correct AngleOffset 
is set for your specific application (refer to the 
SWIFT
®
Mini Transducer Interface product
manual, part number 100-214-316.).
Excessive noise during the 
zero procedure caused an 
incorrect bridge zero, or the 
angle computation to be 
calculated incorrectly.
Electrical Noise: The transducer has power 
conditioning, shielded cables and on-board 
amplification to reduce electrical noise. 
However, if the TI or transducer are very near a 
powerful noise source, some noise can be 
picked up with the signal. 
Check that the TI electronics are grounded 
properly through the power cable to the power 
source. If possible, perform the zeroing and 
shunt calibration procedure away from the noise 
source.
Physical Noise: If the transducer is part of an 
assembly that is connected to active hydraulics 
or other assemblies that may experience 
physical vibrations, these vibrations may be 
picked up as inertial forces read by the 
transducer. Check that the transducer is not 
experiencing dynamic loading or vibrations 
during the zero process or shunt calibration.
Troubleshooting Guide (part 4 of 11)
Symptom
Possible Causes
Solution