MTS SWIFT 50 GLP Sensor Heavy Trucks User Manual
Page 78

SWIFT 50 GLP Sensors
78
Troubleshooting
The shunt reference values 
were changed.
Use the TI2XFER program to upload the 
current calibration file. Check the file to verify 
that the variables ShuntDeltaRef are the same 
as shown on the original calibration file 
provided by MTS. The eight ShuntDeltaMeas 
values should read approximately 0.85 V.
The shunt tolerance is not 
set correctly.
Use the TI2XFER program to upload the 
current calibration file. The ShuntTolerance 
value should be set at 2, indicating an allowable 
shunt variation of 2%. If this value is changed to 
zero or a very small number, the difference 
between the reference and the current shunt 
measurement may fall outside of this tolerance 
and be read as a shunt failure.
Transducer cable problems.
If the ShuntDeltaMeas values are not the same 
as (or within tolerance of the ShuntDeltaRef 
values), note which shunt values are not the 
same.
If all eight are not correct, check the shunt 
cables, signal cable, and power supply.
If four of the signals are not correct, verify that 
both shunt cables are plugged in and are not 
damaged. (Each shunt cable shunts four 
bridges). If connections appear correct, swap 
the transducer cable and check whether the bad 
shunt values follow the cable. This indicates a 
damaged or faulty cable. Check that the cable is 
not damaged, and that connector pins are not 
missing or damaged.
Use the same swapping technique if one or 
more signals is not correct. In this case swap the 
slip ring bracket and note which bridge has the 
different value. If the faulty shunt value stays 
the same with the slip ring bracket, look at the 
connectors for damage or objects plugging the 
connector and contact MTS.
Troubleshooting Guide (part 7 of 11)
Symptom
Possible Causes
Solution
