Ti rear panel, Ti rear panel 41 – MTS SWIFT 40 Sensor User Manual
Page 41

Transducer Interface
SWIFT 40 Sensors
Hardware Overview
41
TI Rear Panel
Transducer Interface Rear Panel
J4 Output connector
The J4 Output connector provides the conditioned sensor outputs that can be
connected to a data acquisition or test control system.
Transducer connector
Connect the data cable from the transducer slip ring to the Transducer Connector.
J2A Shunt A
J2B Shunt B
connectors
The J2 connectors are used for the shunt cables. The shunt cables are connected
to the two 6-pin shunt cal connectors on the front of the transducer.
Comm In
Comm In is an 8-pin modular phone jack connector (RJ-45) that provides a
connection to an RS-232/RS-485 remote host, or to another TI in an RS-485
multidrop communications chain. The cable connected to Comm In selects an
RS-232 or RS-485 interface. The RS-232/RS-485 interfaces on Comm In are
electrically isolated from the TI power. This connection is only needed when you
are downloading new settings to the TI module.
Under normal operating conditions, no host connection is required. The Comm
In and Comm Out connectors allow you to daisy-chain multiple boxes without
Y cables or multiple connections to the host computer.
Comm Out
Comm Out is an 8-pin modular phone jack connector (RJ-45) that provides a
connection to another TI chassis in an RS-485 multidrop communication chain.
J3 Power connector
Connect a power cable from an external 12 V DC source.
Ground Terminals
The ground terminals enable you to ground the TI, and chain several TI boxes
together.
J2A Shunt A
and J2B Shunt B
J4 Output
Comm In and
Comm Out
J3 Power
Ground
Terminals
Transducer
Connector
S20-12