beautypg.com

4 responsibility of the owner, 5 intended use – Memmert Universal ovens U plus, Incubators I plus, Sterilisers S plus User Manual

Page 8

background image

8

D24026 | Date 10/2014

Safety regulations

1.4 Responsibility of the owner

The owner of the appliance

is responsible for the flawless condition of the appliance and for its proper operation in

accordance with its intended use (see page 8);

is responsible for ensuring that persons who are to operate or service the appliance are

qualified to do this, have been instructed accordingly and are familiar with the operating

instructions at hand;

must know about the applicable guidelines, requirements and operational safety regula-

tions, and train staff accordingly;

is responsible for ensuring that unauthorised persons have no access to the appliance;

is responsible for ensuring that the maintenance plan is adhered to and that maintenance

work is carried out properly (see page 55);

has to ensure that the appliance and its surroundings are kept clean and tidy, for example

through corresponding instructions and inspections;

is responsible for ensuring that personal protective clothing is worn by operating person-

nel, e.g. work clothes, safety shoes and protective gloves.

1.5 Intended use

This appliance is exclusively intended for heating up non-explosive substances and objects.

Any other use is improper, and may result in hazards and damage.
The appliance is not explosion-proof (does not comply with the German workplace health &

safety regulation VBG 24). The appliance may only be loaded with materials and substances

which cannot form any toxic or explosive vapours at the set temperature and which cannot

explode, burst or ignite.
The appliance may not be used for drying, vaporising and branding paints or similar materi-

als the solvents of which could form an explosive mixture when combined with air. If there

is any doubt as to the composition of materials, they must not be loaded into the appliance.

Potentially explosive gas-air mixtures must not form, neither in the working chamber nor in

the direct vicinity of the appliance.

Intended use as a medical device

For appliances subject to the 93/42/EEC guideline (Council Directive on the approximation of

the laws of the Member States relating to medical devices), the intended use is defined as

follows:

For appliances of the UF

PLUS

type series: The appliance serves for heating non-sterile cloths

and covers.

For appliances of the IF

PLUS

type series: The appliance serves for heating non-sterile cloths

and covers, as well as for temperature control of rinsing and infusion solutions.

For appliances of the IN

PLUS

type series: The appliance serves for temperature control of

rinsing and infusion solutions.

For appliances of the SF

PLUS

type series: The appliance serves for sterilising medical material

through dry heated air at atmospheric pressure (also see page 54).