Menu functions

7.3.7 Balance

For appliances of the size 55 and above, application-specific correction of the heat output dis-

tribution (balance) between the upper and lower heating groups is possible. The adjustment

range is from –50 % to +50 %.

upper heat output

upper heat output

lower heat output

-20%

lower heat output

+30%

SETUP

Balance

-20

%

2/2

SETUP

Balance

+30

%

2/2

Abb. 25 Heat output distribution (example): The –20 % (left) setting causes the lower heating

groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower

Heat output distribution (example): The –20 % (left) setting causes the lower heating

Heat output distribution (example): The –20 % (left) setting causes the lower heating

heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the

groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower

groups to emit 20 % less heat than the upper ones. The +30 % (right) setting causes the lower

default heat output distribution.

heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the

heating groups to emit 30 % more heat than the upper ones. The 0 % setting restores the

7.3.8 Remote control

In the setup entry

Remote control

, you can set whether the

appliance should be controlled via remote control and if

so, in which mode. These settings are available:

Off

Read Only

Write

Write+Alarm

When the appliance is in remote control mode, the

symbol appears in the temperature display. In the

symbol appears in the temperature display. In the

settings

Write

and

Write+Alarm

, the appliance cannot

be controlled at the ControlCOCKPIT until the remote

control has been switched off (setting Off) or set to

Read

Only

.

In order to use the remote control function, program-

ming skills and special libraries are required.

Setup

Balance

Remote Control

Off

Gateway 192.168.5 .1

+30 %

2/2

23

.2°C

TEMP

Set 180

.0°C