Menu functions

7.3.5 Timer mode

For universal ovens UN

PLUS

/UF

PLUS

PLUS

PLUS

and incubators IN

PLUS

/IF

PLUS

PLUS

PLUS

, you

can choose whether the digital backwards counter with target

time setting (see page 28) should run setpoint-dependent or

not – this determines if the timer should not start until a

tolerance band of ±3 K around the set temperature is reached

( Abb. 24 , B) or if it should start right after activation (A).

t

Timer

t

Timer

A

B

°C/°F

t

6 K

Abb. 24

Timer mode

A

Timer independent of setpoint: Timer starts right after activation

B

Timer setpoint-dependent: Timer does not start until tolerance band is reached

A

Timer independent of setpoint: Timer starts right after activation

A

Timer independent of setpoint: Timer starts right after activation

For sterilisers SN

PLUS

/SF

PLUS

PLUS

PLUS

, the timer is by default setpoint-dependent. To ensure that the

required temperature is maintained for a sufficient period of time, this setting cannot be

changed. If the temperature is no longer within the tolerance band, the sterilisation time

is, for reasons of safety, restarted as soon as the required temperature is reached again.

For universal ovens UN

PLUS

/UF

PLUS

PLUS

PLUS

and incubators IN

PLUS

/IF

PLUS

PLUS

PLUS

, the timer is in this case inter-

rupted and resumed as soon as the required temperature is reached again.

7.3.6 Type of the slide-in unit (Grid or Shelf)

Here, you have to set the type of the slide-in unit (grid or shelf)

used. The selection Shelf enables you to adjust the control

function to the different air flow characteristics in the interior

when using optional sliding shelves instead of the grids that are

part of the standard delivery.

SETUP

Unit

IP address

255.145.136.225

Subnet mask

255.255.0.0

Alarm temp

°C

F

TWW

TWB

Timer mode

Slide-in unit

Grid

Shelf

SETUP

Unit

IP adress

255.145.136.225

Subnet mask

255.255.0.0

°C

°F

Timer mode

Alarm temp

TWW

TWB

Slide-in unit

Grid Shelf