5.5.3 Automatic temperature monitor (ASF)

ASF is a monitoring device that automatically follows the set temperature setpoint within an

adjustable tolerance band (Abb. 20).
The ASF – if switched on – is automatically activated as soon as the actual temperature value

reaches 50 % of the set tolerance band of the setpoint (in the example: 180 °C – 1.5 K) for the

first time (section A).
When the temperature violates the set tolerance band around the setpoint (in the example in

Abb. 20:

180 °C ± 3 K) – e.g. if the door is opened during operation (section B of illustration) – the

alarm is set off. The ASF alarm is automatically triggered as soon as 50 % of the set tolerance

band of the setpoint (in the example: 180 °C ± 1.5 K) are reached again (section C).
If the temperature setpoint is altered, the ASF is automatically disabled temporarily (in this

example: The setpoint is changed from 180 °C to 173 °C, section D) until the tolerance range

of the new temperature setpoint is reached again (section E).

183 °C

177 °C

ASF active

183 °C

177 °C

ASF alarm

ASF active

ASF active

176 °C

170 °C

180 °C

t

A

B

C

D

E

AUTO

AUTO

AUTO

Abb. 20 Schematic diagram of how the ASF temperature monitoring works

5.5.4 Mechanical temperature monitoring: Temperature limiter (TB)

The appliance is equipped with a mechanical temperature limiter (TB) of protection class 1 in

accordance with DIN 12 880.
If the electronic monitoring unit should fail during operation and the factory-set maximum

temperature is exceeded by approx. 20 °C, the temperature limiter, as the final protective

measure, switches off the heating permanently.

5.5.5 Adjusting temperature monitoring

1. Press the activation key to the left of the

ALARM

display. The

min

setting (under-

temperature protection) is automatically

activated.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

180

.0 °C

Set 999

.9 °C

TEMP

TIMER

End 14:45

FLAP

Holz trocknen
aufheizen
09:12h

Fr 20.10.2010 20:31

min

000°C

ALARM

0

12

%rh

°C

DRAPH

off

ON

44

h

:44

m

20

%

ON

max

000

°C

auto off

99K

-

+

ALARM

+-

0.0

K

min

°C

auto

max

120

.0°C

60

0

.