D24026 | Date 10/2014

Putting into operation

4. Putting into operation

Caution:

The first time the appliance is operated, it must not be left unattended until it has reached

the steady state.

4.1 Connecting the appliance

Caution:

Observe the country-specific regulations when making connections (e.g. in Germany DIN

VDE 0100 with residual current circuit breaker). Observe the connection and power rat-

ings (see nameplate and Technical data on page 14). Make sure to establish a safe PE

conductor connection.

Lay the power cable so that

it is always accessible and within reach so it can be disconnected quickly in the event of

failure or emergencies;

no one can trip over it;

it does not come into contact with any hot parts.

230/115-V appliances:

230/115-V appliances:

Plug the provided power cable into the rear of the appli-

ance and connect it to a CEE 7/4 socket (Abb. 10).

400V appliances:

400V appliances:

The power cable is permanently installed. Connect the

plug to a 400 V CEE coupling (Abb. 11).

4.2 Switching on

Switch on the appliance by pressing the On/Off switch

on the front of the appliance ( Abb. 12 ).
The starting process is shown by three animated white

dots

. If the dots have another colour, an error has

. If the dots have another colour, an error has

occurred (see page 38).

After the first start-up, the appliance display is set to

English by default. You can change the language as

described from page 41. However, to get a basic

overview of operating the appliance, you should read

the following chapter first.

Abb. 10 Power connection

230/115 V

Abb. 11 400 V CEE connection

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Abb. 12 Switch on appliance