beautypg.com

Memmert Universal ovens U plus, Incubators I plus, Sterilisers S plus User Manual

Page 27

background image

D24026 | Date 10/2014

27

Operation and control

The status display shows you which operating mode or operating state the appliance is

currently in. The current operating state is highlighted in colour and indicated by the text

display:

Appliance

is in programme mode

Programme is stopped
Appliance is in manual operating mode

The example on the right shows the appliance in man-

ual mode, identified by the coloured hand symbol.

When the appliance is in timer mode,

Timer active

is

displayed:

When the appliance is in remote control mode, the

symbol appears in the temperature display:

5.4.4 Manual mode

In this operating mode, the appliance runs in permanent

operation at the values set on the ControlCOCKPIT.

Adjustment options

Adjustment options

As described in chapter 5.4.2 , you can set the following

parameters after pressing the corresponding activation key (in any sequence):
Temperature

Temperature

Adjustment range: model dependent (see nameplate and technical

date on page 14)

Heating operation is indicated by the symbol.

symbol.

You can select °C or °F as the temperature unit displayed (see

page 43).

24

.4°C

TEMP

180

.4

°C

Set

Air flap position

Air flap position
Adjustment range: 0 % (closed, recirculating operation) to 100 %

(completely opened, fresh air operation) in steps of 10 %

FLAP

%

40

Fan speed

Fan speed

(only for UF/IF/SF

PLUS

appliances)

Adjustment range: 0 to 100 % in steps of 10%

FAN

%

50

Interior lighting (additional option)

Interior lighting (additional option)
Adjustment range: 0 %, 100 %

LIGHT

%

100

Manual Mode

12.Sept.2012 13:44

Timer active

12.Sept.2012 13:44

23

.2°C

TEMP

Set 180

.0°C