beautypg.com

Unp 500, 6 programme mode, Starting a programme – Memmert Universal ovens U plus, Incubators I plus, Sterilisers S plus User Manual

Page 29: Manual mode, Test 012, Ramp 1, Alarm, Graph, Turn the turn control until the, The programme description (in this exam- ple

background image

D24026 | Date 10/2014

29

Operation and control

To deactivate the timer, open the timer display by pressing the

activation key again and then turning the turn control to reduce

the timer setting until --:-- is displayed. Confirm with the confirma-

tion key.

5.4.6 Programme mode

In this operating mode, programmes saved in the appliance can be started with different

combinations of individual parameters (temperature, air flap position, fan speed, working

chamber lighting) at staggered intervals, which the appliance then automatically processes in

sequence. These programmes are not created directly at the appliance but externally at a com-

puter / laptop and using AtmoCONTROL software. Transfer to the appliance is possible using

the provided USB storage medium or via Ethernet.

A description of how to create and save programmes can be found in the separate

AtmoCONTROL software manual.

Starting a programme

Starting a programme

1. Press the activation key to the right of

the status display. The current operating

mode is highlighted automatically, in this

example

Manual Mode

(

).

UNP 500

344

.4

444

4

.

TEMP

Set

°C

°C

LIGHT

100

%

manueller Betrieb

FLAP

40

%

.4

444

4

.

TEMP

Set

°C

°C

FLAP

%

LIGHT

%

TIMER

End

--

--

m

h

- -

- -

- - - - -

Fr 20.10.2010 20:31

ALARM

min

100°C

max

400°C

auto off

+/- 0.0K

0

8

16

40

80

GRAPH

0

8

16

40

80

GRAPH

APH

Manual mode

17:44

13.Sept.2012

Activate

Turn the turn control until the

start

symbol is highlighted. The current

programme is displayed, in this example

Test 012

.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Test 012
ready

10:44

12.Sept.2012

Only the programme currently selected in the menu and shown in the display can be

used. If you want to process another programme, you need to activate it in the menu

first (description from page 48 ).

3. To start the programme, press the confir-

mation key. The programme starts. The

display shows:

the programme description (in this exam-

ple

Test 012

ple

ple

)

the programme segment description, in

this example

Ramp 1

the current run (in case of loops)

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

Test 012
Ramp 1

10:44

12.Sept.2012

You cannot change any parameters (e.g. the temperature) at the appliance while a

programme is running. However, the displays

ALARM

and

GRAPH

can still be used.

TIMER

--

m

--

h

End 9:00 23.11.