beautypg.com

Memmert Universal ovens U plus, Incubators I plus, Sterilisers S plus User Manual

Page 37

background image

D24026 | Date 10/2014

37

Malfunctions, warning and error messages

Description

Cause

Action

See

Temperature alarm and

"TWB" are displayed

TEMP

TWB Set 190.0

°C

195

.4

°C

The electronic

temperature

limiter (TWB) per-

manently switched

off heating.

Deactivate the alarm by pressing

the confirmation key.
Increase the difference between

the monitoring and setpoint

temperature – by either increas-

ing the max value of the temper-

ature monitoring or decreasing

the setpoint temperature.
If the alarm continues: Contact

customer service

page

33

page 2

Temperature alarm and

"TB" are displayed

TEMP

TB

230

.4

°C

The mechanical

temperature lim-

iter (TB) perma-

nently switched

off heating.

Switch off the appliance and

leave to cool down. Contact

customer service and have the

error rectified (e.g. by replacing

the temperature sensor).

page 2

6.2 Malfunctions, operating problems and appliance errors

Error description

Cause of error

Rectifying errors

See

Displays are dark

External power supply

was interrupted

Check the

power supply

page

22

Miniature fuse, appliance

fuse or power module

faulty

Contact customer

service

page 2

Displays cannot be

activated

Appliance locked by user ID Unlock with user ID

page

53

The appliance is in pro-

gramme, timer or remote

control mode (mode "Write"

or "Write + Alarm")

Wait until the end of

the programme or

timer mode or switch

off the remote control

Displays suddenly look

different

Appliance is in "wrong"

mode

Change to operating or

menu mode by pressing

the

MENU

key