49

Menu mode

7.5 Calibration

To guarantee perfect control, we recommend to calibrate the appliance once a year.

7.5.1 Temperature adjustment

The appliances are temperature calibrated and adjusted at the factory. In case readjustment

should be necessary later on – for example due to influence of the chamber load – the ap-

pliance can be calibrated customer-specifically using three calibration temperatures of your

choice:

Cal1

Temperature calibration at low temperature

Cal2

Temperature calibration at medium temperature

Cal3

Temperature calibration at high temperature

For temperature adjustment, you will need a calibrated reference measuring device.

CAL 1

+0,5°C

CAL 3

+1,6°C

CAL 2

-0,4°C

Factory calibration

20°C

30°C

40°C

0°C

10°C

Fig. 26

Schematic example of temperature adjustment

Example: Temperature deviation at 30 °C should be corrected.

1. Press the activation key to the right of the

CALIB

display. The display is enlarged

and the temperature adjustment option

is automatically highlighted.

UFP 100

Cal1

°C

-0,2

K

Cal2

100.0

°C

+0,1

K

Cal3

180.0

°C

-0,2

K

JUSTIEREN

Temperatur

PROGRAM

TIME

USER ID

PROTOCOL

PROGRAM

TIME

USER ID

PROTOCOL

letzte Justierung 12.10.2012 12:00

40.0

C

K

0,2

-

C

K

0,1

+

C

K

0,2

-

Calibration

Last updated 12.10.2012 12:00:00

Cal1

Cal2

Cal3

20.0

37.0

5.0

Temperature

Humidity

2. Press the confirmation key repeatedly,

until the calibration temperature Cal2 is

selected.

Hauptschalter

>

Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,

ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung

getroffen hat.

Die hier gezeigt Variante, ist die von uns empfohlene Richtung.

UF 110

PLUS

ON

344

.4

444

4

.

TEMP

Set

°C

°C

FLAP

40

%

TIMER

h

d

10

02

End

Sept.

29

22 24

FAN

%

0

%

100

ALARM

max

444

.4°C

auto

+/-

0.0

K

min

444

.4°C

O

O

ON

N

N

N

O

O

N

O

Manu

44.Sept

C

K

0,2

-

C

K

0,1

+

C

K

0,2

-

JUSTIEREN

Temperatur

letzte Kalibrierung 12.10.2012 12:00

Cal1

Cal2

Cal3

37.0

5.0

20.0