5 calibration, Ufp 100, 1 temperature adjustment – Memmert ICH/ ICH L Climate chambers User Manual
Page 49: Calib, Temperatur, Menu mode, Cal1 temperature calibration at low temperature, Cal2 temperature calibration at medium temperature, Justieren
D30381 | Date 10/2014
49
Menu mode
7.5 Calibration
To guarantee perfect control, we recommend to calibrate the appliance once a year.
7.5.1 Temperature adjustment
The appliances are temperature calibrated and adjusted at the factory. In case readjustment
should be necessary later on – for example due to influence of the chamber load – the ap-
pliance can be calibrated customer-specifically using three calibration temperatures of your
choice:
►
Cal1
Temperature calibration at low temperature
►
Cal2
Temperature calibration at medium temperature
►
Cal3
Temperature calibration at high temperature
For temperature adjustment, you will need a calibrated reference measuring device.
CAL 1
+0,5°C
CAL 3
+1,6°C
CAL 2
-0,4°C
Factory calibration
20°C
30°C
40°C
0°C
10°C
Fig. 26
Schematic example of temperature adjustment
Example: Temperature deviation at 30 °C should be corrected.
1. Press the activation key to the right of the
CALIB
display. The display is enlarged
and the temperature adjustment option
is automatically highlighted.
UFP 100
Cal1
°C
-0,2
K
Cal2
100.0
°C
+0,1
K
Cal3
180.0
°C
-0,2
K
JUSTIEREN
Temperatur
PROGRAM
TIME
USER ID
PROTOCOL
PROGRAM
TIME
USER ID
PROTOCOL
letzte Justierung 12.10.2012 12:00
40.0
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
Calibration
Last updated 12.10.2012 12:00:00
Cal1
Cal2
Cal3
20.0
37.0
5.0
Temperature
Humidity
2. Press the confirmation key repeatedly,
until the calibration temperature Cal2 is
selected.
Hauptschalter
>
Zu der Typenbezeichnung gibt es momentan drei Entwurfsrichtungen,
ich kann Ihnen noch nicht sagen, ob Memmert hierzu schon eine Entscheidung
getroffen hat.
Die hier gezeigt Variante, ist die von uns empfohlene Richtung.
UF 110
PLUS
ON
344
.4
444
4
.
TEMP
Set
°C
°C
FLAP
40
%
TIMER
h
d
10
02
End
Sept.
29
22 24
FAN
%
0
%
100
ALARM
max
444
.4°C
auto
+/-
0.0
K
min
444
.4°C
O
O
ON
N
N
N
O
O
N
O
Manu
44.Sept
C
K
0,2
-
C
K
0,1
+
C
K
0,2
-
JUSTIEREN
Temperatur
letzte Kalibrierung 12.10.2012 12:00
Cal1
Cal2
Cal3
37.0
5.0
20.0