beautypg.com

Memmert ICH/ ICH L Climate chambers User Manual

Page 40

background image

40

D30381 | Date 10/2014

Malfunctions, warning and error messages

6.2 Malfunctions, operating problems and appliance errors

Error description

Cause of errors

Rectifying errors

See

Displays are dark

External power supply

was interrupted

Check the

power supply

page 21

Miniature fuse, appliance

fuse or power module

faulty

Contact customer

service

page 2

Displays cannot be

activated

Appliance locked by USER

ID

Unlock with USER ID

page 56

The appliance is in pro-

gramme, timer or remote

control mode (mode "Write"

or "Write + Alarm")

Wait until the end of

the programme or

timer mode or switch

off the remote control

Displays suddenly look

different

Appliance is in "wrong"

mode

Change to operating

or menu mode by

pressing the

MENU

key

Display T:E-3 in the

temperature display

TEMP

T:E-3

Set 37.0

°C

37

.4

°C

Temperature operating sen-

sor defective. The monitor-

ing sensor takes over the

measurement function.

The appliance can

be temporarily

operated

Contact customer

service as soon as

possible

page 2

Error message AI E-3 in

the temperature display

TEMP

AI E-3

Set 37.0

°C

37

.4

°C

Temperature monitoring

sensor defective. The oper-

ating sensor takes over the

measurement function.

The appliance can

temporarily be

kept in service

Contact customer

service as soon as

possible

page 2

Error message E-3 in

the temperature display

TEMP

Set 45.0

°C

E-3

°C

Operating and monitoring

sensor defective

Switch off appli-

ance.

Remove the cham-

ber load

Contact customer

service

page 2