beautypg.com

3 power failure – Memmert ICH/ ICH L Climate chambers User Manual

Page 41

background image

D30381 | Date 10/2014

41

Malfunctions, warning and error messages

Error description

Cause of errors

Rectifying errors

See

Error message E-6 in

the humidity display

HUMIDITY

Set 50

.0%rh

%rh

E-6

Humidity sensor defective

No humidity con-

trol possible

Contact customer

service

page 2

When switching on

the appliance, the start

animation is displayed

in another colour than

white

Cyan

:

Not enough storage

space on the SD card

Red

:

The system files could

not be loaded

Orange

: The

fonts and images could

not be loaded

Contact customer

service

Contact customer

service

Download the firm-

ware update at mem-

mert.com and install it

page 2

page 2

6.3 Power failure

In case of a power failure, the appliance operates as follows:

In manual mode

After power supply has been restored, operation is continued with the parameters set. The

time and duration of the power failure are documented in the log memory.

In timer or programme mode

In timer or programme mode

In case of an interruption of the power supply of less than 60 minutes, the current pro-

gramme is continued from the point at which it was interrupted. For longer interruptions of

the power supply, all appliance functions (heating, fan etc.) are switched off.

In remote control mode:

The previous values are restored. If a programme has been initiated via remote control, it is

continued.