Delivery, transport and setting up
3.6 Setting up
Warning!
Due to its centre of gravity, the appliance can fall over to the front
and injure you or other people. Always attach the appliance to a wall
with the tilt protection (see page 20). In case there is not enough
space, do not put the appliance into operation and do not open the
door. Contact the Memmert service (see page 2).
The appliance may only be installed on the fl oor.
3.6.1 Preconditions
The installation site must be flat and horizontal and must be able to reliably bear the weight
of the appliance (see Technical data on page 14). Do not place the appliance on a flamma-
ble surface.
Depending on the model (see nameplate), a 230 V or 115 V power connection must be avail-
able at the installation site.
The distance between the wall and the rear of the appliance must be at least 15 cm. The clear-
ance from the ceiling must not be less than 20 cm and the side clearance from walls or nearby
appliances must not be less than 5 cm (Fig. 9). Sufficient air circulation in the vicinity of the
appliance must be guaranteed at all times. Do not put anything in front of the fan opening of
the cooling unit at the appliance front.
For appliances with castors, these need to be positioned in forward direction at all times.
≥ 5 cm
≥ 15 cm
≥ 20 cm
≥ 5 cm
Fig. 9
Minimum clearance from walls and ceiling