beautypg.com

Maintenance and service, 1 cleaning – Memmert HPP, Peltier-cooled incubators IPP plus Constant climate chambers User Manual

Page 59

background image

D24030 | Date 10/2014

59

Maintenance and service

8. Maintenance and service

8.1 Cleaning

Warning!

Danger due to electric shock. Before doing any maintenance

work, pull out the mains plug.

Warning!

In case of appliances of a certain size, you can get accidentally

locked in, which is life-threatening. Do not climb into the appliance!

Caution!

Danger of cuts due to sharp edges. Always wear gloves when wor-

king in the chamber interior.

8.1.1 Working chamber and metal surfaces

Regular cleaning of the easy-to-clean working chamber prevents build up of material remains

that could impair the appearance and functionality of the stainless steel working chamber

over time.
The metal surfaces of the appliance can be cleaned with normal stainless steel cleaning

agents. Make sure that no rusty objects come into contact with the working chamber or with

the stainless steel housing. Rust deposits can lead to an infection of the stainless steel. If rust

spots should appear on the surface of the working chamber due to impurities, the affected

area must be immediately cleaned and polished.

8.1.2 Plastic parts

Do not clean the ControlCOCKPIT and other plastic parts of the appliance with caustic or

solvent-based cleaning agents.

8.1.3 Glass surfaces

Glass surfaces can be cleaned with a commercially available glass cleaner.

8.1.4 Peltier cooling module

In order to guarantee perfect

function and long lifetime of the

Peltier cooling modules, it is

absolutely essential that you

remove dust deposits from the

heat sink on the rear of the

appliance (with a vacuum

cleaner, paintbrush or bottle

brush, depending on the

amount).
To make cleaning easier, the

cover can be removed after the

screws have been loosened (Fig.

30).

Fig. 30

Cover of the Peltier cooling modules on the

rear of the appliance