Malfunctions, warning and error messages

6.1.2 Humidity monitoring (for constant climate chambers HPP only)

Description

Cause

Action

See

Error symbol

HUMIDITY

Set 55

.0%rh

%rh

55

.4

Water tank

empty

Fill the water tank with deminera-

lised/distilled water and press the

confirmation key

page

22

Alarm display (

MaxAl

)

HUMIDITY

MaxAl

Set 70

.0%rh

%rh

75

.4

Upper humidity

limit exceeded

Open the door for 30 sec. and

wait to see if the appliance reliab-

ly adjusts to the setpoint.
If the error occurs again, contact

customer service.

page 2

Alarm display (

MinAl

)

HUMIDITY

MinAl

Set 60

.0%rh

%rh

55

.4

Humidity below

lower limit

Check if the door is closed.
Check the water supply and the

filling level of the water tank. If

required, refill the water tank.
If the error occurs again, contact

customer service.

page

22

page 2

6.2 Malfunctions, operating problems and appliance errors

Error description

Cause of error

Rectifying errors

See

Displays are dark

External power supply

was interrupted

Check the power supply page

22

Miniature fuse, appliance

fuse or power module

faulty

Contact customer

service

page 2

Displays cannot be

activated

Appliance locked by user ID Unlock with user ID

page

58

The appliance is in pro-

gramme, timer or remote

control mode (mode "Write"

or "Write + Alarm")

Wait until the end of

the programme or

timer mode or switch

off the remote control

Displays suddenly look

different

Appliance is in "wrong"

mode

Change to operating or

menu mode by pressing

the

MENU

key