9 applied directives and standards, 10 declaration of conformity, 11 ambient conditions – Memmert HPP, Peltier-cooled incubators IPP plus Constant climate chambers User Manual
Page 15

D24030 | Date 10/2014
15
Construction and description
2.9 Applied directives and standards
►
Directive 2004/108/EC amended (Directive of the council on harmonisation of the laws of
the member states on electromagnetic compatibility). Fulfilled standards:
DIN EN 61326:2004-05, EN 61326:1997, EN 61326/A1:1998, EN 61326/A2:2001
EN 61326/A2:2003
►
Directive 2006/95/EC amended (Directive of the council on harmonisation of the laws of
member states relating to electrical equipment designed for use within certain voltage
limits). Standards complied with:
DIN EN 61 010-1 (VDE 0411 Part 1):2002-08, DIN EN 61 010-2-010 (VDE 0411 Part
2-010):2004-06, EN 61 010-1:2001, EN 61 010-2-010:2003
2.10 Declaration of conformity
You can download the EC declaration of conformity of the appliance online:
English: http://www.memmert.com/en/service/downloads/ce-statement/
German: http://www.memmert.com/de/service/downloads/eg-konformitaetserklaerung/
2.11 Ambient conditions
►
The appliance may only be used in enclosed rooms and under the following ambient
conditions:
Ambient temperature
16 ºC to 40 ºC
Humidity rh
max. 70 %, non-condensing
Overvoltage category
II
Pollution degree
2
Altitude of installation
max. 2,000 m above sea level
►
The appliance may not be used in areas where there is a risk of explosion. The ambient air
must not contain any explosive dusts, gases, vapours or gas-air mixtures. The appliance is
not explosion-proof.
►
Heavy dust production or aggressive vapours in the vicinity of the appliance could lead to
sedimentation in the interior and, as a consequence, could result in short circuits or dam-
age to electrical parts. For this reason, sufficient measures to prevent large clouds of dust
or aggressive vapours from developing should be taken.