Construction and description

Range A:

Range A:

In this range, temperature and humidity can be combined as you please, without resulting in

any significant condensation. If there are extreme ambient conditions, the working range may

be restricted.

Range B:

Range B:

If the specified range is exceeded upwards, e.g. 80 % rh at 60°C, the hot steam fed in will im-

mediately condense, due to the dew-point, at the coldest point in the appliance.

Range C:

Range C:

At low temperatures and low relative air humidity, the effective range is heavily dependent on

the degree of humidity of the chamber load.

2.4 Material

For the outer housing, MEMMERT deploys stainless steel (Mat.No. 1.4016 – ASTM 430) and

for the interior, stainless steel (Mat.No. 1.4301 – ASTM 304) is used, which stands out through

its high stability, optimal hygienic properties and corrosion-resistance towards many (but not

all!) chemical compounds (caution for example with chlorine compounds).
The chamber load for the appliance must be carefully checked for chemical compatibility with

the materials mentioned. A material resistance table can be requested from the manufacturer.

2.5 Electrical equipment

Operating voltage and current consumption: See nameplate

Protection class I, i.e. operating insulation with PE conductor in accordance with EN 61010

Protection type IP 20 acc. to EN 60 529

Interference suppression acc. to EN 55011 class B

Appliance fuse: Fusible link 250 V/15 A quick-blow

The temperature controller is protected with a miniature fuse 100 mA (160 mA at 115 V)

2.6 Connections and interfaces

2.6.1 Electrical connection

This appliance is intended for operation on an electrical power system with a system imped-

ance Z

max

of a maximum of 0.292 ohm at the point of transfer (service line). The operator

max

max

must ensure that the appliance is operated only on an electrical power system that meets

these requirements. If necessary, you can ask your local energy supply company what the

system impedance is.
Observe the country-specific regulations when connecting (e.g. in Germany DIN VDE 0100

with residual current circuit breaker).